Title

DETERMINING TG-43 BRACHYTHERAPY DOSIMETRY PARAMETERSAND DOSE DISTRIBUTION FOR A 131CS SOURCE MODEL CS-1

Pages

  85-90

Abstract

 Background: Monte Carlo determination of TG-43 brachytherapy dosimetry parameters and DOSE DISTRIBUTION calculation for 131CSsource model CS-1 are presented in this study. Materials and Methods: The DOSE DISTRIBUTION was calculated around the 131CS Model CS-1 located in the center of 30 cm x30 cm x30 cm water, and soft tissue phantoms cube using MCNP CODE by Monte Carlo method. The percentage depth dose (POD) variation along the different axis, parallel and perpendicular, the source was calculated. Then, the ISODOSE CURVES for 100%, 75%, 50% and 25% POD were constructed. Finally, F(r,q) and g(r) dosimetry parameters of TG-43 protocol have been determined. Results: Results obtained show that the Monte Carlo method could only calculate dose deposition in high gradient region, near the source, accurately. The energy cut off was found to be 1 eV and the error in the calculations was less than 2%. Conclusion: The ISODOSE CURVES of the CS-1 131CS source were constructed from dose calculation by MCNPcode. The calculated dosimetry parameters for the source were in agreement with previously published results.

Cites

  • No record.
  • References

  • No record.
  • Cite

    APA: Copy

    YAZDANI, M., & MOULAVI, A.A.. (2007). DETERMINING TG-43 BRACHYTHERAPY DOSIMETRY PARAMETERSAND DOSE DISTRIBUTION FOR A 131CS SOURCE MODEL CS-1. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 5(2), 85-90. SID. https://sid.ir/paper/286418/en

    Vancouver: Copy

    YAZDANI M., MOULAVI A.A.. DETERMINING TG-43 BRACHYTHERAPY DOSIMETRY PARAMETERSAND DOSE DISTRIBUTION FOR A 131CS SOURCE MODEL CS-1. INTERNATIONAL JOURNAL OF RADIATION RESEARCH[Internet]. 2007;5(2):85-90. Available from: https://sid.ir/paper/286418/en

    IEEE: Copy

    M. YAZDANI, and A.A. MOULAVI, “DETERMINING TG-43 BRACHYTHERAPY DOSIMETRY PARAMETERSAND DOSE DISTRIBUTION FOR A 131CS SOURCE MODEL CS-1,” INTERNATIONAL JOURNAL OF RADIATION RESEARCH, vol. 5, no. 2, pp. 85–90, 2007, [Online]. Available: https://sid.ir/paper/286418/en

    Related Journal Papers

  • No record.
  • Related Seminar Papers

  • No record.
  • Related Plans

  • No record.
  • Recommended Workshops